3 calibrating (manual), Calibrating (manual) – Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual

Page 36

Advertising
background image

Calibration in control standard (determining the cell constant)

WQL-Cond

132

ba75822e01

03/2011

6.3

Calibrating (manual)

Preparatory

activities

Calibration

Note
We recommend to always calibrate with the protective hood screwed on so
the measuring cell is protected. Use a stand as necessary.

Note
For manual calibration with a control standard, the temperature dependent
conductivity values of the standard are required. Normally a table with these
values is enclosed with the control standard.

z

Prior to manual calibration, please switch off the temperature
compensation (menu, Device / Set).

z

Determine the conductivity of the used standard solution using the
temperature measured by the logger during calibration (interpolate the
table values as necessary).

z

After the calibration is completed, restore the required setting for the
temperature compensation.

1

Connect the WQL-Cond logger to a USB port of your PC with the
USB cable.

2

Keep a control standard of your choice ready.

3

In the Device menu, click the menu item, Calibrate.
The calibration window appears.

4

Immerse the logger in the control standard.