79

Adjustmentandcommissioning

8.10.23. setting the parameters of a ph/redox module

Refer to chap. 8.9 to access Parameters menu.

if "Temp. calib" = "Manual"

This is
when the
device is be-
ing parame-
tered............
....................

This is
when the
device is be-
ing parame-
tered............
....................

Mx:pH/ORP

Temp.Adjust

RTD

Temp. calib

Parameters

PT100

None

PT1000

ENTERING

Calibration limits

Warning low:

Warning high:

Error low:

pH zero

Temperature

Auto

Manual

Auto

Manual

Buffer

Hamilton

DIN19267

Error high:

ENTERING
ENTERING
ENTERING
ENTERING

Warning low:

Warning high:

Error low:

pH slope

Error high:

ENTERING
ENTERING
ENTERING
ENTERING

Warning low:

Warning high:

Error low:

ORP Offset

Error high:

ENTERING
ENTERING
ENTERING
ENTERING

°C

ENTERING

°C

ENTERING

if "Temperature" = "Manual"

if "Temperature" = "Auto"

RTD: Choose the type of temperature sensor connected to the module.
TEMPERATURE: Choose the value of the temperature used in the process (particularly to compensate for the pH
measurement):

• choose "Auto": the temperature of the fluid is measured by the sensor.

• choose "Manual": enter the value of the process temperature (in °C) in the next field, e.g. when no temperature

sensor is connected to the module.

TEMP. ADJUST: The temperature measured can be corrected by an offset value. Enter the offset value in °C.
TEMP. CALIB: Choose the value of the temperature used when calibrating the probe/sensor:
• choose "Auto": the temperature of the fluid is measured by the sensor.

• choose "Manual": enter the value of the calibration temperature (in °C) in the next field, e.g. when no temper-

ature sensor is connected to the module.

BUFFER: Choose the type of buffer solution used for automatic calibration of the pH probe/sensor, "Hamilton"
solutions sold by Bürkert or solution conforming to DIN 19267:

• the multiCELL automatically recognizes the following "Hamilton" pH buffers: 2 ; 4,01 ; 7 ; 10 ; 12 ;

• the multiCELL automatically recognizes the following "DIN19267" pH buffers: 1,09 ; 4,65 ; 6,79 ; 9,23 ; 12,75.

Type 8619

english