Modifying the calibration menu access code, Adjusting the current outputs – Burkert Type 8619 User Manual

Page 86

Advertising
background image

84

Adjustmentandcommissioning

To deactivate the HOLD mode:

access the "HOLD" function

choose "Disable"

confirm by "OK".

8.11.2. modifying the calibration menu access code

Refer to chap. 8.9 to access Calibration menu. If the default access code "0000" is kept, the device does not
request it to access the Calibration menu.

Enter the new
CALIBRATION menu
access code

Confirm the new
code

Calibration

System

Code

0***

Confirm code

0***

8.11.3. Adjusting the current outputs

Ensure that Hold mode is deactivated before adjusting the current outputs: the

icon is displayed.

Refer to chap. 8.9 to access Calibration menu.

Calibration

M0:Outputs

AO1/AO2

4mA

20mA

ENTERING

ENTERING

Mx:Outputs

4mA: Adjust the offset of the current output.
When the "4mA" function is selected, the multiCELL generates a current of 4 mA:

measure the current emitted by the 4-20 mA output using a multimeter;

enter the value given by the multimeter.

20mA: Adjust the span on current output 1 or current output 2.
When the "20mA" function is selected, the multiCELL generates a current of 20 mA:

measure the current emitted by the 4-20 mA output using a multimeter;

enter the value given by the multimeter.

Type 8619

english

Advertising