Burkert Type 8619 User Manual

Page 95

Advertising
background image

93

Adjustmentandcommissioning

The multiCELL displays the calibration result.

1)

Rinse
7.001 pH
-59.6 mV
25.00 °C

Immerse the clean probe in the first
buffer solution: the multiCELL displays
the measured pH of the solution.

Enter the pH of the buffer solution (indicated on the bottle)

Wait until the measured pH stabilises.

Calibration

Mx:pH/ORP

pH manual calib.

OK

F4

1st point

7.035 pH
7.000 pH

pH Calibr. result

Zero: 7.000 pH
Slope:-59.15mV/pH

Calibrate a 2nd

point?

NO

Rinse the probe.

confirm rinsing by "OK" when the pH
reaches the desired value.

OK

F4

2nd point

2.967 pH
2.000 pH

Immerse the clean probe in the second buffer solution:
the multiCELL displays the measured pH of the solution.

Enter the pH of the second buffer solution (indicated on the
bottle)

Wait until the measured pH stabilises.

OK

F4

pH Calibr. result

Zero: 7.000 pH
Slope:-59.15mV/pH

YES

The multiCELL displays the
calibration result.

1)

1)

• a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.

• a possible "error" message indicates that the probe must be replaced.

Type 8619

english