Adjustmentandcommissioning

Immerse the cleancell in the reference solution: the multiCELL displays the
measured conductivity value of the solution

Enter the conductivity value of the ref-
erence solution used (indicated on the
bottle).

Modify the unit, if necessary.

Calibration

Cond. manual calib

Mx:conductivity

OK

F4

Manual calibration
5.023 µS/cm
5.000 µS/cm

Calibration result
Cell constant:
1.00000 /cm

The multiCELL displays the calibration result.

8.12. "Diagnostics" menu

8.12.1. modifying the "Diagnostics" menu access code

Refer to chap. 8.9 to access the Diagnostics menu. If the default access code "0000" is kept, the device does
not request it to access the Diagnostics menu.

Enter the new
DIAGNOSTICS
menu access code

Confirm the new
code

Diagnostics

System

Code

0***

Confirm code

0***

8.12.2. monitoring the current or voltage value received on the

analogue inputs

This function is used to define the behaviour of the limits entered by the user are reached.

Refer to chap. 8.9 to access the Diagnostics menu.