Burkert Type 8619 User Manual

Page 89

Advertising
background image

87

Adjustmentandcommissioning

Connect the analogue input AI1 or AI2 to the current or voltage analogue output of an external instrument (for
instance, the 4-20 mA output of a pressure measuring device type 8311).

The analogue input signal is adjusted (slope and offset) and the multiCELL
displays the input calibrated value.

Generate a 4 mA current on the 4-20 mA output of the instrument con-
nected to the analogue input.

Enter 4 mA.

Wait for the measurement to stabilize.

Calibration

Mx:Inputs

Current calibr.

OK

F4

1st point

4.001 mA
4.000 mA

Calibrate using a

2nd point?

NO

2nd point

20.001 mA
20.000 mA

Generate a 20 mA current on the 4-20 mA output of the instrument con-
nected to the analogue input.

Enter 20 mA.

Wait for the measurement to stabilize.

OK

F4

YES

The multiCELL adjusts the
offset.

AI1/AI2

Value received on the analogue input.

Calibration result

16.02 mA

SAVE

F4

Figure 69 :

Example for the calibration of an analogue input with respect to the current output of a pressure measuring
device type 8311

Type 8619