Calibration menu, English – Burkert Type 8175 User Manual

Page 105

Advertising
background image

37

8175

ENGLISH

CALIBRATION MENU

ENTER

ENTER

ENTER

0......9

4.4.6 Teach-In procedures

The Teach-In function identifies and fixes the relationship between the transmitter display
and the real distance, level or volume to ensure accurate measurement.

The transmitter automatically selects the appropriate Teach-In procedure
(distance, level or volume) depending on the configuration and units se-
lected within the engineering units function - section 4.4.2

-

For the Teach-In procedures it is advisable to select either filter 0 or 3 (depending on

echo filtering) during the Teach-In procedure for quick calculation of the measure and

after select the desired filter.

-

If volume has been selected for measurement and gas characteristics have to be

defined, follow the procedures below:

-

Select distance or level with appropriate units within the engineering units.

-

Complete a 2 point Teach-In for identification of the gas characteristics and ‘END‘ the

menu to save the parameters.

-

Return to the engineering units and select volume and the appropriate units.

-

Complete a volume Teach-In to fix the transmitter display and the real distance, level or

volume

measurement.

4.4.6.1 Teach-In for level and distance

This function allows the relationship between the transmitter and fluid to be defined by
distance or level.

-

For distance measurement it is not necessary to complete a Teach-In procedure

although for accurate measurement it is advised to complete at least a 1 point Teach-In.

-

If the transmitter has been configured as level, it is necessary to complete at least a 1

point

Teach-In.

CURRENT

1 or 2 point Level or distance
Teach-In procedure

No Teach-In chosen
and advance to the next
option.

Reset the distance reference
point

(Reset is not displayed if configured as level)

TEACH-IN

TEACH N

TEACH Y

RESET

A

B