8

8175

ENGLISH

QUICKSTART

To access the CALIBRATION MENU simutaneously press

ENTER

for 5 seconds.

-

The ‘Enter‘ key can be locked to avoid accidental or unauthorised access - section 3.3.

-

Reference can additionally be made to the menu guide - section 4.2.

-

It is important to exit the menu calibration after entering the desired distance units and

re-enter the menu to configure the volume units.

2.2.3

Programming - Volume

The actions which are highlighted in grey

must be fully completed for accurate

measurement.

Choose the

desired language

Select the desired

distance units

Quit this menu and

re-enter

Select the desired

volume units

Select the desired

filter

Enter the

Sound velocity

Perform a Teach-In

procedure

Enter the current

output range

Configure relays

1 & 2 ( if available )

See section 4.4.1

See section 4.4.2

See section 4.4.2

See section 4.4.3

See section 4.4.4

See section 4.4.6

See section 4.4.7

See section 4.4.8

Measurement or Simulation (optional)

See the next page.

It is important to quit this
calibration menu and re-
enter after entering the

distance units to ensure
correct measurement.

Choose filters 0-2 if no echo filtering
of obstructing obstacles is required
and filters 3-9 if echo filtering is
required.

331.4 m/s within the air with a devia-
tion of 0.59m/s/°C.

For ex. 4mA -> 1.2m

3

and 20mA -> 10.6m

3