Specifications, English, 1 specifications – Burkert Type 8175 User Manual

Page 126

Advertising
background image

58

8175

ENGLISH

SPECIFICATIONS

6.1 SPECIFICATIONS

Specifications in relation to the process

Measurement
Measurement type

Level, distance or volume measurement

Measurement range

0.30 to 10 meters compact version and

0.30 to 7 meters wall and panel versions

Measurement range in the
following conditions:

Measure of the fluid without foam,

Tambient = 20°C ; Tfluid = 20°C

Ambient pressure value : atmopheric pressure

Accuracy

+/- 0.25% of the full scale

+/- 0.15% after calibration with Teach-In

Resolution

+/-

3mm

Full beam width

8 degrees conical

Pulse rate

8 pulses per second

Blocking distance

30 cm from the base of the sensor


Installation
Pressure rating

2 bar at 25°C maximum

Mounting threads

G 2‘‘ or NPT 2‘‘

Fluid temperature

-40 to +80°C / -40 to +176°F

Gasket material

FPM or EPDM (option)

Specifications in relation to the outputs

Electrical connection
Power

supply

18-32VDC

115/230 VAC 50-60Hz (+/- 10% VAC only)

Consumption

200mA

maximum


Proportional ouput
Output type

Current output of 4-20mA (22mA error signal)

Accuracy

+ / - 2%

Wiring

Sinking or sourcing

Maximum load

1300 Ohms at 32V

1000 Ohms at 24V

550 Ohms at 18V

Output calibration

4-20mA or inverted to 20-4mA adjustable

Relay ouput
Output type

2 relays (3A), normally open

Load

DC : 250V, 3A

AC : 250V, 3A

Life expectancy

100 000 cycles (minimum)

Switching

Thresholds, hysteresis and delay programmable

Advertising