3.1 GENERAL INSTALLATION GUIDELINES

Prior to the installation of the 8175 ultrasonic level transmitter, a clear mounting location for
the product must be determined.

-

The level transmitter or sensor 8175 / 8170 are designed for liquid

level measurement. It is the user‘s responsability to test the

functionality of the device for any other material such as powders,

granulas

etc.

-

The 8175 ultrasonic level transmitter or sensor must be installed

perpendicular to the process medium.

-

Always protect the transmitter from the rain, ultraviolet radiations

and electromagnetic perturbations.

Tanks with bulged or cone-shaped roofs

Mounting on a shaft

Beam cone

min. 30 cm

H

R/2

D

R

D (m)

R (cm)

1 30
2 60
3 80
4 100
5 110
6 120
7 130
8 135
9 140
10

140

8

Do not install the transmitter at the center of a
bulged or cone-shaped roof but at half-radius
from the center.

H: see table hereafter.

For the mounting on a shaft:
- respect the max. shaft length:

- burr the inside of the shaft.

The ultrasonic pulses are emitted from the
base of the transmitter with an 8° beam angle.
Suppress all the obstacles inside the beam
cone to avoid false echos.

D

H

D

H

D [mm]

65

80

100

H [mm]

100

250

400