Luminex 100 IS Version 2.1 User Manual

Page 17

Advertising
background image

x

MAP Technology

Luminex xMAP Technology

PN 89-00002-00-062 Rev. A

11

Figure 2. New Advanced Batch—General Tab

2. Fill in the Name, Description, and Operator fields.

3. Edit the Sample Size field if desired. The default value is 50 uL.

This field accepts values from 20 to 200.

4. Edit the DD Gate field as desired. The default value is 7500 to

15000. The field accepts a range of 0 to 32767.

5. Edit the Timeout field as desired. The default value is 25

seconds. The field accepts a range of 0 to 400 where 0 is no
timeout.

6. Click the menu bar next to 100 regions to change this default

value. The available bead maps are 25, 50, and 100 regions.

7. Click the Bead Set tab. The Options Bead Set tab dialog appears.

Select the checkboxes next to each desired Bead Set for this
batch. Click the Select All button to select all the listed bead sets.
Click the Deselect All button to deselect any selected bead sets.

Note:

The Luminex 100 IS

software uses only the 100
region Bead Map. While in
Acquisition Detail, you can
view only the data (for 25, 50,
and 100 regions) and save to
a csv file.

Advertising