Luminex 100 IS Version 2.1 User Manual

Page 36

Advertising
background image

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP Technology

30

PN 89-00002-00-062 Rev. A

The Apply Suffix button applies the number one to the first test
name. For each successive test name it applies the test name and
number at one increment higher than the previous test name. For
example: Suffix = Protein, then Test 1 is Protein 1, Test 2 is
Protein 2, and so on.

The Apply Unit button applies the same unit value that you
entered on the first line to all other tests. For example, if you
enter the mg/uL value for the first Test Unit, then select the
Apply Unit button to make all test units mg/µL.

7. Enter the Number of Standards (assay) for this product and press

Enter key.

8. Enter the Reagent Name for each standard in this product. Each

standard reagent must test for every analyte (bead set) used in the
assay.

9. Enter the Number of Controls (assay) for this product and press

Enter key.

10. Enter the Reagent Name for each control in this product. Each

control reagent must test for every analyte (bead set) used in the
assay.

11. After entering all the information, check it carefully. When all

the information is correct, click Save to save the product.

12. Click Close.

Note:

If the reagent names are

similar, click the Apply Suffix
button to ensure that each
name is unique.