Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP

Technology

74

PN 89-00002-00-062 Rev. A

2. Disperse the pellet with sonication, and vortex the container for

20 seconds.

3. Dispense 1.0

× 10

5

Avidin-modified microspheres into a 1.5 mL

microcentrifuge tube.

4. Centrifuge the microspheres for 1 minute at

≥ 8,000 × g and

remove the supernatant.

5. Add 250 µL of the REACTION BUFFER (See Technical

note 2). Vortex.

6. Add 250 µL of the Biotin-conjugated molecule solution to the

microsphere suspension. Vortex immediately.

7. Rotate the mixture for 30 minutes in the dark at room

temperature.

8. Centrifuge the microspheres for 1 minute at

≥ 8,000 × g.

Aspirate the supernatant.

9. Add 500 µL of BLOCKING/STORAGE BUFFER to the

microspheres and vortex. Centrifuge the tube for 60 seconds.
Aspirate the supernatant.

10. Repeat step 9.

11. Resuspend the microspheres in the desired volume of

BLOCKING/STORAGE BUFFER.

12. Enumerate the microsphere preparation.

13. Store the preparation at 2

o

C - 8°C. Protect it from light.

Technical Notes

1. The concentration of Biotin-conjugated molecules is provided as

a guide for assay development. The concentration required for
optimal reactivity must be determined empirically for each
specific biotin-conjugated molecule.

2. The reaction volume can vary with the reagent volume. If the

reagent is limited, perform the reaction in a proportionately low
volume.