Luminex 100 IS Version 2.1 User Manual

Page 79

Advertising
background image

PN 89-00002-00-062 Rev. A

73

x

MAP

Technology

3. The optimal coupling concentration for a given oligonucleotide

is determined by coupling at various concentrations within the
recommended range of 1 to 50 µM.

4. Minimize the exposure of EDC to air; secure closures on stock

and dry aliquot containers, store desiccated at -20°C. Use
aliquots immediately and discard containers after use. Make a
fresh 10 mg/mL EDC solution before each addition.

Binding Biotin-conjugated Molecules to xMAP

LumAvidin-Modified Microspheres

Introduction

Use the protocols included here as a general starting point for
developing assays. All assays should be optimized for your reagents
in your specific application. You will discover the methods that give
you the best results by starting with these guidelines and modifying
them for your specific needs.

Equipment

Vortex

Micropipetters (1 µL - 1000 µL)

Microcentrifuge

Timer

Rotator

Materials

xMAP LumAvidin-modified microspheres—LIMIT
EXPOSURE TO LIGHT!

Biotin-conjugated molecule

Microcentrifuge tubes: 1.5 mL, polypropylene

REACTION BUFFER:
Phosphate Buffered Saline (pH 7.3 ± 0.1), Bovine Serum
Albumin (10 mg/mL)

BLOCKING/STORAGE BUFFER:
Phosphate Buffered Saline (pH 7.3 ± 0.1), Tween 20
(0.02% v/v), Bovine Serum Albumin (1 mg/mL), Sodium Azide
(0.05% w/v)

Preparation

Dilute the Biotin-conjugated molecule in REACTION BUFFER to a
concentration of 4 - 4000 nM (See Technical note 1).

Procedure

1. Centrifuge the Avidin-modified microsphere stock for 1 minute

at

≥ 8,000 × g.

Advertising