Introduction, Equipment, Materials – Luminex 100 IS Version 2.1 User Manual

Page 74

Advertising
background image

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP

Technology

68

PN 89-00002-00-062 Rev. A

Two-Step Carbodiimide Coupling of Protein to xMAP

Carboxylated Microspheres

Introduction

Use the protocols included here as a general starting point for
developing assays. All assays should be optimized for your reagents
in your specific application. You will discover the methods that give
you the best results by starting with these guidelines and modifying
them for your specific needs.

Equipment

Vortex

Sonicator bath

Micropipetters (1 µL - 1000 µL)

Microcentrifuge

Analytical balance

Timer

Rotator

Materials

xMAP carboxylated microspheres—LIMIT EXPOSURE TO
LIGHT!

Microcentrifuge tubes: 1.5 mL, polypropylene (See Technical
note 4)

ACTIVATION BUFFER:
0.1 M Sodium Phosphate (pH 6.2 ± 0.2)

COUPLING BUFFER:
50 mM MES (pH 5.0)

WASH BUFFER:
Phosphate Buffered Saline (pH 7.4 ± 0.1), Tween

®

20

(0.05% v/v)

BLOCKING/STORAGE BUFFER:
Phosphate Buffered Saline (pH 7.4 ± 0.1), Bovine Serum
Albumin (10 mg/mL), Sodium Azide (0.05% w/v)

Sulfo-NHS:
N-Hydroxysulfosuccinimide sodium salt, Pierce Chemicals

EDC:
1-Ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride,
Pierce Chemicals

Protein for coupling (See Technical note 2)