Performance specifications, Speed, Accuracy and precision – Luminex 100 IS Version 2.1 User Manual

Page 8: Sensitivity

Advertising
background image

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP Technology

2

PN 89-00002-00-062 Rev. A

Performance

Specifications

Speed

Speed: greater than or equal to 933 MHz Pentium III processor
with 256 MB RAM

USB cabling for fast data transfer

Automatic transfer of assay templates and new reagent
information into the system using a 3½” inch floppy disk or large
capacity read/write CD

Installation < 4 hours

System calibration < 10 minutes

System controls < 10 minutes

Barcode reader entry of sample IDs

Automatic post analysis

Analyze one 96-well plate/hour depending on manufacturer’s kit

Up to 100 xMAP microsphere sets per sample

Sheath flowrate 90

µL/sec. ± 5 µL

Sample injection rate into detector area 1

µL/sec. ± 0.05 µL

System warmup 30 minutes: Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.

Accuracy and Precision

Sample uptake volume accuracy ± 5%

Classification of xMAP microspheres > 80.0%

Misclassification of xMAP microspheres < 0.5%

Temperature control 0°C to ± 2°C of target

Internal sample carry over < 0.9%

Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity value

Sensitivity

Detect 1000 fluorochromes Phycoerythrin (PE) per microsphere

Reporter channel dynamic range: 3.5 decades