Plate layout tab – Luminex xPONENT 3.1 Rev 2 (IVD) User Manual

Page 28

Advertising
background image

1/2 Log - Creates a 1:3.16 dilution, or half of each 1:10 (Log) from each previous

iteration.

• Alternatively, type a number for your own dilution factor.

Apply Dilution - Applies the dilution selected in the Dilution list.

NOTE: Click a column header to resort the display.

NOTE: Click the Reagent column header to resort the order from the

highest number standard to standard number one. This is useful for
applying dilutions in which the last standard is the highest standard.

Assay Control Information - Lists the selected control reagents. The list displays the
Reagent, Name, Lot Number, Expiration, and Manufacturer. Existing control lot
information can be applied or new information can be typed manually.

Apply Ctrl Lot - Opens the Select Lot dialog box. Select a lot from the list and then click

OK.

Show Value - Expected, Low, and High set the expected, lowest, or highest acceptable

concentration of the analyte in the sample.

Apply Values - Applies a value down or across the list of analytes.

Cancel - Returns to the Batches tab.

Back - Returns to the previous tab.

Next - Opens the Plate Layout tab.

Plate Layout Tab

Use this tab to define commands that apply to one or more wells. You can define off-plate
and maintenance commands.

xPONENT

®

3.1 Rev. 2 Software User Manual

18

For In Vitro Diagnostic Use

Advertising