Error conditions – Agilent Technologies L4400 User Manual

Page 109

Advertising
background image

Operating and Programming

3

L4400 User’s Guide

97

Error Conditions

Usage:

All L4400 Series Instruments

When the L4400 instrument’s front panel ATTN LED is red, one or more
command syntax or hardware errors have been detected. A record of up
to 20 errors can be stored in the instrument’s error queue.

For a complete listing of the error messages, see the Programmer’s
Reference
Help file located on the Agilent L4400 Product Reference
CD- ROM that ships with the instrument.

A special global error queue holds all power- on and hardware- related

errors (e.g., over- temperature, Safety Interlock, etc.).

Errors are retrieved in first- in- first- out (FIFO) order. The first error

returned is the first error that was stored. Errors are cleared as you
read them. Once you have read all of the interface- specific errors, the
errors in the global queue are retrieved.

Errors are cleared as you read them. When you have read all errors

from the interface- specific and global error queues, the ATTN
annunciator turns off.

If more than 20 errors have occurred, the last error stored in the

queue (the most recent error) is replaced with - 350,“Error queue
overflow”. No additional errors are stored until you remove errors
from the queue. If no errors have occurred when you read the error
queue, the instrument responds with +0,“No error”.

The interface- specific and global error queues are cleared by the

*CLS

(Clear Status) command and when power is cycled. The errors are
also cleared when you read the error queue. The error queue is not
cleared by a Factory Reset (

*RST

command) or an Instrument Preset

(

SYSTem:PRESet

command).

The following command is used to read errors from the error queue. One
error is read per command.

SYSTem:ERRor?