One-wire mode – Agilent Technologies L4400 User Manual

Page 131

Advertising
background image

L4433A Dual/Quad 4x8 Reed Matrix

5

L4400 User’s Guide

119

One-Wire Mode

To physically configure the module in 1- wire mode, use the 34933T- 002
terminal block,

or a compatible standard or custom cable. If using a

standard or custom cable, make sure you connect interlock pins 17 and
33 on the Matrix 2 D- sub connector. Refer to the pinout drawing and
table on

page 125

.

In 1- wire mode, the L4433A module contains four matrices (1 through
4), each with 32 1- wire crosspoint non- latching reed relays organized in
a 4- row by 8- column configuration. Every row and column has one wire
each. Each crosspoint relay has a unique channel number representing
the matrix, and the single- wire row and column that intersect to make
the crosspoint. For example, channel 218 represents Matrix 2, row 1 and
column 8. See the simplified schematic on

page 124

.

In 1- wire mode, you can close no more than 40 channels simultaneously
due to power dissipation. For example, with one analog bus relay closed
you can close up to a maximum of 39 channel relays. If you try to close
more than the allowed number of channels, you will receive an error
message.

You can connect any combination of inputs and outputs at the same
time. However, only Matrix 3 and Matrix 4 in 1- wire mode of this module
connect to the analog buses. By closing channels 921 and 922 you can
connect row 1 and row 2 respectively to the HI (H) and LO (L) lines of
ABus1 and ABus2.

You can connect multiple matrix modules externally and/or through the
analog buses for applications that require large matrices. For information
on linking multiple matrix modules, refer to

page 116

of this chapter.

When the power is off, matrix relays and analog bus relays open.