Totalizer Mode

Totalizer mode is the default operating mode for the counters. When the
counter is configured for

TOTalizer

mode, it automatically starts running.

The totalized count can be read, reset, scanned, and monitored.

The simplest way to take a totalizer measurement is to use the

MEASure

form of the command. For example, the following command configures the
totalizer on the first bank, initiates the measurement, and returns the
result. The data is returned in a floating point format.

MEAS:COUN:TOT? READ, (@1301)

You can also reset the totalizer count by setting the parameter to

RRESet

.

For example, the following command configures the totalizer on the first
bank, initiates the measurement, and returns the result. The totalize count
is reset when the data is read.

MEAS:COUN:TOT? RRES, (@1301)

Totalizer counts begin as soon as the channel is configured for the totalize
measurement. You can stop a count by sending

SENSe:COUNter:ABORt

command and restart the count using the

SENSe:COUNter:INITiate

command.

The slope of the edges being counted can be configured using the

SENSe:COUNter:SLOPe

command. By default, when started, the totalizer

counts rising edges.

Additionally, you can control when the edges are counted by setting the
gate source to external and providing a gate signal on the gate input.

In external gate mode the counter totalizes when the gate is asserted.

The gate time setting controls how long the counter totalizes. Once the
external gate has been de- asserted a new measurement must be armed via
the

SENSe:COUNter:INITiate

command. The figure below shows an

externally gated totalizer measurement. The number of totalized counts is
‘5’ in this particular example.