Agilent Technologies L4400 User Manual

Page 90

Advertising
background image

78

L4400 User’s Guide

3

Operating and Programming

Figure 3-12. Channel Delay.

You can set the channel delay to any value between 0 seconds and

60 seconds, with 1 ms resolution. You can select a different delay for
each channel.

You can select a unique delay for every channel on the module.

The channel delay is valid only while scanning. If no channels have

been assigned to the scan list, the specified channel delay is ignored
(no error is generated).

The default channel delay is 0.0 seconds.

A Factory Reset (

*RST

command) sets the channel delay to 0.0s.

The following command adds a 2- second channel delay to the specified
channels.

ROUT:CHAN:DELAY 2,(@1003,1013)

t

Scan List

t

Channel Delay

(0 to 60 seconds)

Ch 1

Ch 3

Ch 4

Ch 2

Ch 5

Ch 6

t

1

t

3

t

4

t

6

t

2

t

5

Advertising