Ext trig/alarms/dio port, Gpib connector, Lan port – Agilent Technologies L4400 User Manual

Page 63

Advertising
background image

Operating and Programming

3

L4400 User’s Guide

51

See “Scanning with External Instruments” later in this chapter for
information on how the analog bus is used for scanning a channel list
with an external DMM.

Ext Trig/Alarms/DIO Port

The external trigger, alarms, and DIO port enables you to synchronize
scanning between a switching instrument such as the L4421A and an
external DMM. The port also allows you to output alarm signals to an
external device or control system. Figure 3- 4 shows the pin out and
signal definitions for the port.

Input

5 V

0 V

> 1 µs

Output

3.3 V

0 V

Approx. 2 µs

External Trigger Usage

Alarm Usage

5

1

9

6

Alarm 1 output (pin 1)

Alarm 2 output (pin 2)

Gnd (pin 9)

Channel closed output

(Trig Out - pin 5)

Gnd (pin 9)

Channel advance input

(Trig In - pin 6)

5

1

9

6

or

Figure 3-4. External Trigger and Alarm Port Pin Definitions.

GPIB Connector

The GPIB interface is available on all L4400 series instruments as
Option- GPIB. This option must be purchased with the product. Products
not ordered with the GPIB interface cannot be reconfigured to add it later.

LAN Port

The LAN port on the L4400 series instruments supports 10 Mbps and
100 Mbps data transfer rates (10BaseT/100BaseTx). The port is Non
Auto- MDIX which means that the LAN crossover cable supplied with the