General purpose switch instrument – Agilent Technologies L4400 User Manual

Page 140

Advertising
background image

128

L4400 User’s Guide

6

L4437A General Purpose Switch

General Purpose Switch Instrument

The L4437A General Purpose (GP) switch can be used to route signals or
control other system devices.

The L4437A provides independent control of 32 latching relays:

Twenty- eight Form C relays rated for 1 A at 60 W per channel

Four Form A relays rated for 5 A at 150 W per channel.

The L4437A contains armature- latching relays, and you can use the
switch for device actuation, digital output, or combine it with additional
switch instruments to create flexible switching topologies. You can close
multiple channels at the same time. The L4437A does not implement an
analog bus.

A temperature sensor on the instrument triggers system interrupts when
high- carry current- induced heat on the instrument is excessive. This
over- temperature situation generates an SRQ event when the factory- set
70

o

C threshold is reached. It is up to the user to determine what, if any,

action should be taken.

Reactive loads (those that include significant inductance or capacitance)
can cause voltage spikes or current spikes during switching operations.
The L4437A is designed for switching reactive loads. The optional 34937T
terminal block has solder pads for adding snubber circuits for the 5 A
relays to reduce the reactive transients. See the drawings on

page 134

for the locations of snubber circuit pads and installation information
about a snubber circuit.

A hardware jumper on the L4437A allows you to define the power- failure
states for the instrument’s 5 A latching relays. Depending on the position
of the jumper, the 5 A relays will either open or maintain state when
system power failure occurs. When shipped from the factory, the
power- fail jumper is in “MAINTAIN” position (all relays maintain their
present state when power fails).

Remove the L4437A instrument sub- assembly from the instrument carrier
and then remove the sheet metal cover from the sub- assembly. Move the
position of the jumper mounted on the sub- assembly. See the figure
below for the jumper’s location.

WA R N I N G

Before changing the position of the jumper, turn off the
instrument and remove all external connections. Wait five to ten
seconds to allow the instrument’s internal capacitors to
discharge.

Advertising