L4452A Multifunction Module with DIO, D/A, and Totalizer

Example: Configuring the totalizer for count This command configures the
totalizer to count on the rising edge or falling edge of the input signal.
The following command configures the totalizer (channel 5) to count on
the negative edge (falling) of the input signal.

TOTalize:SLOPe NEGative,(@1005)

Example: Clearing count on the totalizer channel This command immediately
clears the count on the specified totalizer channel (channel 5).

TOTalize:CLEAR:IMMediate (@1005)

DAC Output

Example: Setting output voltage This command sets the output voltage level
for the specified DAC channels. The following command outputs +2.5 V
DC on DAC channels 6 and 7.

SOURce:VOLTage 2.5,(@1006,1007)

Querying and Resetting the L4452A

Example: Querying the instrument identify The following command returns
the identify of the multifunction module.

SYSTem:CTYPe? 1

Example: Resetting the instrument to its power-on state The following
command resets the multifunction module to its power- on state.

SYSTem:CPON 1