Scan Trigger Source

Usage:

L4421A 40- Channel Armature Multiplexer

L4450A 64- Bit Digital I/O (digital input, counter channels only)

L4452A Multifunction Module (digital input, totalizer channels only)

You can configure the event or action that controls the onset of each
sweep through the scan list (a sweep is one pass through the scan list):

You can set the instrument’s internal timer to automatically scan at a

specific interval. You can also program a time delay between channels
in the scan list (see

“Channel Delay"

on page 77).

You can start a scan when an external TTL trigger pulse is received.

You can start a scan when an alarm event (L4450A, L4452A) is logged

on the channel being monitored.

Interval Scanning

In this configuration, you control the frequency of scan sweeps by
selecting a wait period from the start of one trigger to the start of the
next trigger (called the trigger- to- trigger interval). If the scan interval is
less than the time required to measure all channels in the scan list, the
instrument will scan continuously, as fast as possible (no error is
generated).

Figure 3-10. Trigger-to-Trigger Interval.

You can set the scan interval to any value between 0 seconds and

99:59:59 hours (359,999 seconds), with 1 ms resolution.

Once you have initiated the scan, the instrument will continue

scanning until you stop it or until the trigger count is reached. See

“Trigger Count"

on page 75 for more information.

t

Sweep 1

Trigger 1

Sweep 2

Sweep

n

Trigger 2

. . .

Trigger Timer

(0 to 359,999 seconds)