Scanning on alarm – Agilent Technologies L4400 User Manual

Page 86

Advertising
background image

74

L4400 User’s Guide

3

Operating and Programming

Scanning on Alarm

Usage:

L4450A 64- Bit Digital I/O (counter channels only)

L4452A Multifunction Module (totalizer channels only)

In this configuration, the instrument initiates a scan each time a reading
crosses an alarm limit on a channel. You can also assign alarms to
channels on the digital modules. For example, you can generate an alarm
when a specific bit pattern or bit pattern change is detected on a digital
input channel or when a specific count is reached on a totalizer channel.

For complete details on configuring and using alarms, refer to

“Alarm Limits"

on page 84.

In this scan configuration, you can use the Monitor function to

continuously take readings on a selected channel and wait for an
alarm on that channel. Channels do not have to be part of an active
scan list to be monitored; however, the channel must be configured for
a measurement in order to be monitored.

All readings from the scan are stored in the instrument’s volatile

memory.

Readings accumulate in memory until the scan is terminated

(until the trigger count is reached or until you abort the scan).

You can specify a trigger count which sets the number of scan trigger

commands that will be accepted before terminating the scan. See

“Trigger Count"

on page 75 for more information.

Mx+B scaling and alarm limits are applied to measurements during a

manual scanning operation and all data is stored in volatile memory.

The following program segment configures the instrument to continuously
scan when an alarm is detected.

TRIG:SOURCE ALARM1

TRIG:SOURCE:ALARM CONT


CALC:LIM:UPPER 10.25,(@1003)

CALC:LIM:UPPER:STATE ON,(@1003)

OUTPUT:ALARM1:SOURCE (@1003)


ROUT:MON:CHAN (@1003)

ROUT:MON:CHAN:ENABLE ON,(@1003)

ROUT:MON:STATE ON


INIT

Select alarm configuration

Select continuous scan mode

Set upper alarm limit

Enable alarms

Report alarms on Alarm 1

Select monitor channel

Enable monitoring on channel

Enable monitor mode

Initiate the scan

N O T E