background image

116

L4400 User’s Guide

5

L4433A Dual/Quad 4x8 Reed Matrix

Example: Querying channels for open or close state The following command
returns a 1 (true) or 0 (false) state of channel 204.

ROUTe:CLOSe (@1204)

ROUTe:CLOSe? (@1204) !Returns a 1

ROUTe:OPEN? (@1204) !Returns a 0

Example: Querying the system for module Identify The following command
returns the identity of the L4433A:

SYSTem:CTYPe? 1

For the L4433A, the query response may include a suffix to
indicate a 1-wire configuration. For example, the response for the
L4433A will be either "L4433A" (2-wire mode) or
"L4433A-1W" (1-wire mode).

Reading Cycle Count and Resetting the Power-On State

Example: Reading the cycle count for a relay The following command returns
the cycle count on channels 304 and 308.

DIAGnostic:RELay:CYCLes? (@1304,1308)

Example: Resetting the power-on state The following command sets the
L4433A to its power- on state.

SYSTem:CPON 1

Linking Multiple L4433A Instruments

You can link multiple L4433A instruments to form a larger matrix. The
following two drawings show two- module connections through rows and
columns.

You can connect rows in separate instruments using external wiring. Or,
using Bank 2 matrices, you can connect through the analog busses. For a
clear idea of how matrices are arranged and their connections to the
analog busses, see the simplified schematics later in this chapter. You
must use external wiring whenever you connect:

rows in Matrix 1 of separate modules

rows in Matrix 1 to rows in Matrix 2 on the same or separate modules

columns of two matrices on the same or separate modules

N O T E