Continuous drive current, Using continuous drive, Using single drive switches and attenuators – Agilent Technologies L4400 User Manual

Page 158

Advertising
background image

146

L4400 User’s Guide

7

L4445A Microwave Switch/Attenuator Driver

Using Single Drive Switches and Attenuators

Some microwave

switches require a single drive. With single drive devices

the channel numbering is not consecutive across all channels in a bank
(refer to the channel numbering description on

page 143

).

The L4445A can provide single drive devices with either pulsed or
continuous drive current. Settings and parameters for continuous drive
mode are given in the next section.

Continuous Drive Current

Driving non- latching devices requires a power supply capable of handling
sustained high current requirements. You may only use continuous drive
current with channel configured for single drive. Additionally, to prevent
power supply loading, care must be taken when operating more than one
continuous drive at a time. The actual drive may be configured as either
TTL or open- collector operation.

Using Continuous Drive

The diagram below illustrates the continuous drive signals for two
channels (switches) and the relationship of the drive parameters to the
power supply requirements.

T(Setttle)

T(Recovery)

Drive Ch 1

Drive Ch2

Start Drive
Channel 1

Start Drive
Channel 2

Channel 1 Position
Indicators Evaluated

As shown in the diagram, the drive signal is initially applied to channel 1.
Drive is applied to channel 2 only after a power supply recovery period has
elapsed T(Recovery). The power supply recovery time is set using the the

ROUTe:CHANnel:DRIVe:TIMe:RECovery

command. This parameter may be

set individually for each channel or will default to 0.0 ms following either a

SYSTem:RMODule:RESet

or

ROUTe:RMODule:BANK:PRESet

command.

If you are verifying the channel closure (see

page 149

), you may also

specify a T(Settle) parameter. This parameter ensures the switch has had
time to change state before the position indicator is evaluated. This
parameter may be set individually for each channel or will default to 0.0
ms following either a

SYSTem:RMODule:PRESet

or

ROUTe:RMODule:BANK:PRESet

command.