Instrument considerations, Environmental operating conditions – Agilent Technologies L4400 User Manual

Page 14

Advertising
background image

2

L4400 User’s Guide

1

Introduction to the L4400 Series LXI Instruments

Instrument Considerations

This section lists important items and actions that can affect the
operation of your modules.

Environmental Operating Conditions

The L4400 Series LXI modules are designed to operate in a temperature
range of 0 °C to +55 °C with non- condensing humidity. The maximum
humidity is 80% at 40 °C or higher. Do not use in locations where
conductive dust or electrolytic salt dust may be present.

The modules should be operated in an indoor environment where
temperature and humidity are controlled. Condensation can pose a
potential shock hazard. Condensation can occur when the modules are
moved from a cold to a warm environment, or if the temperature and/or
humidity of the environment changes quickly.

The following table shows maximum voltage ratings for each module.
If conditions change, ensure that condensation has evaporated and the
instrument has thermally stabilized until pollution degree 1 conditions
are restored before turning on power to the equipment.

Table 1-1. L4400 Series LXI Instrument Voltage Ratings.

Instrument

Pollution Degree 1 Specifications

Pollution Degree 2 Specifications

L4421A

40 channels, 300V rms or DC, 1A,

60 VA/channel

40 channels, 100V rms or DC, 1A,

60 VA/channel

L4433A

Dual/quad 4x8 matrix, 150 Vpeak,

0.5A, 10 VA/channel

L4437A

L4445A

See Chapter 7 - L4445A

See Chapter 7 - L4445A

L4450A

64 channels, 5V, 30 mA Max

64 channels, 5v, 30 mA Max

L4451A

4 channels 16V, 20 mA

4 channels, 16V, 20 mA

L4452A

32 DIO channels, 42V, 400 mA,

2 channel DAC, 12V, 10 mA

32 DIO channels, 42V, 400 mA,

2 channel DAC, 12V, 10 mA

Dual/quad 4x8 matrix, 100 Vpeak,

0.5 A, 10 VA per channel

28 channels, 300 V rms or DC, 1A,

60 VA per channel
4 channels, 250 V rms or 30 VDC,

5A, 150 VA per channel

28 channels, 100 V rms or DC, 1A,

60 VA per channel
4 channels, 100 V rms or 30 VDC,

5A, 150 VA per channel

Advertising