L4445A Microwave Switch/Attenuator Driver

7

L4400 User’s Guide

189

Example: Configuring a Single Drive Channel

The following example illustrates the sequence of commands to configure a
single drive channel with continuous drive. In the example, the operations
are directed to channel 1 on remote module 3.

The drive source must be disabled before configuring pulse or paired
modes. The channel is then un- paired and the pulse mode disabled
(enables continuous drive). Power supply recovery time and settling time

is then set to 10 ms and 12 ms, respectively. Verify is then enabled.

The switches are set to a

CLOSe

default state and

OCOLlector

drive with

an

EXTernal

power supply is selected. The channel is closed. The final

query of the channel state involves querying both verified state and
whether channel drive is occurring.

ROUTe:RMODule:DRIVe:SOURce OFF,(@1300)

ROUTe:CHANnel:DRIVe:PAIRed OFF,(@1301)

ROUTe:CHANnel:DRIVe:PULSe:MODE OFF,(@1301)

ROUTe:CHANnel:DRIVe:TIME:SETTle 0.010,(@1301)

ROUTe:CHANnel:DRIVe:TIME:RECovery 0.012,(@1301)

ROUTe:CHANnel:VERify ON,(@1301)

ROUTe:CHANnel:DRIVe:CLOSe:DEFault (@1301)

ROUTe:RMOD:BANK:DRIVe:MODE OCOLlector,BANK1,(@1300)

ROUTe:RMODule:DRIVe:SOURce EXT,(@1300)


ROUT:CLOSe (@1301)

ROUT:CLOSe? (@1301)

ROUTe:CHANnel:DRIVE:STATe? (@1301)

The

ROUTe:CHANnel:DRIVE:STATe?

query returns a 0 if the channel is not

being driven and a 1 if the channel is being driven.