Manual scanning – Agilent Technologies L4400 User Manual

Page 85

Advertising
background image

Operating and Programming

3

L4400 User’s Guide

73

Mx+B scaling and alarm limits (L4450A, L4452A) are applied to

measurements during a scan and all data is stored in volatile memory.

The

CONFigure

and

MEASure?

commands (L4450A, L4452A)

automatically set the scan interval to immediate (0 seconds) and the
scan count to 1 sweep.

The instrument sets the scan interval to immediate (0 seconds) after a

Factory Reset (

*RST

command). An Instrument Preset (

SYSTem:PRESet

command) or Card Reset (

SYSTem:CPON

command) does not change

the setting.

The following program segment configures the instrument for an interval
scan.

TRIG:SOURCE TIMER

TRIG:TIMER 5

TRIG:COUNT 2

INIT

Select interval time mode

Set the scan interval to 5 seconds

Sweep the scan list 2 times

Initiate the scan

Note: To stop a scan, send the

ABORt

command.

Manual Scanning

In this configuration, the instrument waits for a command before
sweeping through the scan list.

All readings from the scan are stored in volatile memory.

Readings accumulate in memory until the scan is terminated

(until the trigger count is reached or until you abort the scan).

You can specify a trigger count which sets the number of scan trigger

commands that will be accepted before terminating the scan. See

“Trigger Count"

on page 75 for more information.

Mx+B scaling and alarm limits (L4450A, L4452A) are applied to

measurements during a manual scanning operation and all data is
stored in volatile memory.

The following program segment configures the instrument for a manual
scanning operation.

TRIG:SOURCE BUS

TRIG:COUNT 2

INIT

Select bus (manual) mode

Sweep the scan list 2 times

Initiate the scan

Then, send the

*TRG

(trigger) command to begin each scan sweep.

Note: To stop a scan, send the

ABORt

command.