1.4.3 The NDIR Module Building blocks

Infrared Source
A current-regulated infrared (IR) source provides a photon stream in

the range of 2 to 5 microns. Light is collimated and directed through

the sample cell to the optical block. The source is modulated at 1 Hz,

with a duty cycle of 50%.








Figure 1.4 The Optics of the NDIR Module.

Sample Cell

The sample cell is a gold-coated glass chamber that provides a

means for the infrared source to pass through the sample gas.

Infrared-transparent sapphire windows are provided at each end of

the cell. A microprocessor controlled transducer and thermistor

provide measurement compensation for sample gas temperature and

pressure variances. Sample gas exits the cell at the end opposite the

IR source and is

directed to the UA1372 manifold for O

2

or O

2

/NO

measurement

.

Optical Block

Infrared light not absorbed by the sample gas is transmitted to the

optical assembly. The remaining light is directed to the band pass

filters and on through to the quad detector.
The filter wavelengths are optimised to match the absorption spectra

of Hydrocarbons, Carbon Dioxide and Carbon Monoxide.

The filter associated with Hydrocarbons is centred on 3.46 µm (2889

cm

-1

), the filter associated with Carbon Dioxide is centred on 4.43

IR

SOURCE

SAMPLE CELL

BAND
PASS
FILTERS


ELECTRONICS