LumaSense Technologies INNOVA 1316A-3 User Manual

Page 54

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 54 of 96

Figure 4.1. Status Window displaying “Zero required”.


During a zero point calibration the sample gas is taken through the Zero

inlet on the rear panel of the monitor with a flow rate of 700 ml/min.

ambient air is used for the zero calibration.

Zero Request


The Zero Request (see

figure 4.1

) is set true under the following

conditions:
At POR (power on / reset). When the Multi Gas Monitor is turned ON or

reset, all gas channels will temporarily report 0 gas concentrations (volume

% or ppm). Gas concentration data is only valid after a successful zero

calibration.
After transitioning from standby to normal operating mode.
Zero request interval timeout (90 seconds after POR, followed by a 3

minute then a 6 minute interval and 30 minutes thereafter).
After a ±3°C change in detector temperature.
When an ADC input range error occurs on one or more IR channel (CO

2

,

CO, Ethanol, reference).

Advertising