1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 41 of 96

Stop Soft Key

On pressing the Stop Soft Key data transmission is stopped.

Stopping a measurement sequence can also be accomplished by

pulling down the Bench Command menu and select Data/Status.

Using Stop Continuous in the Data Status picture (See figure 3.6)

data transmission is stopped.

Zero Soft Key


On pressing the Zero Soft Key a zero calibration is performed on the

NDIR Module and if an Oxygen sensor is mounted the Oxygen sensor

are span calibrated with room air.

Gas concentration data is only valid after a successful zero

calibration.

Zero request interval timeout (90 seconds after POR (power-on-

reset), followed by a 3 minute then a 6 minute interval and 30

minutes thereafter).

Zero is also done when:

¨

After a ±3°C change in detector temperature.

¨

When an ADC input range error occurs on one or more IR

channel (CO

2

, CO, Ethanol, reference).

Zero calibration compensates for conditions that affect gas

concentration measurement.

Examples of these conditions are as follows:


Changes in ambient temperature.

Short-term sample cell contamination caused by inadequate particle

filtering.

Long-term sample cell contamination caused by deposits on the

sample cell wall or windows.


Zero calibration execution time is typically 30 to 45 seconds,

including a purge time that depends on the pneumatics configuration

and a fixed calibration time of 20 seconds.