6 operating modes of the multi gas monitor 1316a-3, 1 start-up operating mode – LumaSense Technologies INNOVA 1316A-3 User Manual

Page 49

Advertising
background image

Chapter 3
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 49 of 96

Figure 3.11: The Soft Reset Window.

Press the Send Button and the Instrument is Reset and ready for set-

up.

3.6 Operating Modes of the Multi Gas Monitor

1316A-3

The Multi Gas Monitor 1316A-3’s operations comprise four modes of

operation.

¨

Start-up

¨

Normal

¨

Standby

¨

System Fault

3.6.1 Start-up Operating Mode

After POR (power on / reset), the Multi Gas Monitor 1316A-3

performs the following:

¨

Start-up and initialization.

¨

Self-test.

¨

Thermal warm-up. After a cold start or a Software Reset

bench command the warm-up time period is 35 seconds

maximum. After leaving standby mode the typical warm-

Advertising