2.1 Installing the Monitor

The rear panel has a serial-link and a USB socket for the PC

interface. The mains input, sample gas inlet and outlet for expelled

gas samples are located on the rear panel.

2.1.1 Operating Environment

The following environmental conditions must be observed:

·

Safe ambient operating temperatures range from 5°C to 40°C

·

Humidity should not exceed 90% RH non-condensing.

Warning! The monitor is not intended for use in explosive

environments.

2.1.2 The Gas Sample Line

The gas to be sampled (measured) is drawn from the sample point

via a sample line at a flow rate of 150 ml/min. Various types of

sample lines are available depending on the individual application.

The sample line tubing between the sample point and the monitor

should be kept as short as possible to assist response times. If you

are unsure about the type of sample line for your application, contact

your LumaSense representative for advice before installing the

system.

Up to 100 m tube (AF0614) with internal diameter 3mm can be

attached. The response time is then 70 sec.

Note: You must use a filter in the gas-line immediately before the

gas-line enters the monitor. The UD 5091A in-line filter (optional

accessory) is ideal for this purpose.

The gas inlet for the sample line is located on the rear panel of the

monitor. If the dew point of the gas sample is comparable to the

temperature in the monitor, an inline Genie membrane separator

UA1365 can be used to remove water droplets and eliminate the risk

of condensation inside the monitor

2.1.3 Mains Supply

Warning. The monitor must be used with protective earth.

The monitor operates in the AC voltage range from 100 – 240 V on a

single phase supply, 50 Hz – 60 Hz, 40 VA.

The power cable is connected to a socket on the rear panel. The on-

off switch is located immediately above this socket.