4.3.7 To perform a span calibration with multi component gas mixture.

If more than one gas is to be span calibrated at the same time this

can be performed using multi component gas mixtures.

In the Span calibration window (see

figure 4.22

) select the gases to

be calibrated and key in the appropriate concentrations from the

certificate from the gas cylinder.

Before doing the Span calibration please make sure that Zero

calibration is set to “Manual” and the mode is set to 3 as described in

section 4.3.5

.

1. Use

to tick which of the CO, CO

2

, Ethanol gasses that you

want to be calibrated.

2. Type in the correct concentrations for the ticked CO, CO

2

,

Ethanol gasses.

3. Connect the calibration-gas bottle as described in

section

4.3.4

.

4. Open the gas valve and adjust the flow so that a small surplus

flow exists in the side-stream tube.

5 Press the Send button to start the gain calibration.

After completion of the span calibration routine:

The validity of each new span calibration constant (CO

2

, CO,

Ethanol) is checked.
New span calibration constants for the Ethanol, CO and CO

2

channels should not significantly differ from the factory span

calibration constants.
If a newly calculated span calibration constant does differ by

greater than ±30% from the original factory value, the

appropriate span fail status (CO

2

Span Fail, CO Span Fail, Ethanol

Span Fail), is set true see figure 4.34.: