4.3.4 Connecting the Calibration Gas

Warning! Never make a direct connection from the pressure valve

on a gas cylinder. Always use a side-stream connection like the one

shown in

figure 4.12

.

20090630_1316

Surplus flow

approx. 200 ml/min

800 ml/min

Multi Gas Monitor - INNOVA 1316

UD 5091

Gas Inlet

Cal. Gas

Figure 4.12 Typical side-stream connection set-up used during calibration

4.3.5 Span Calibration

Turn ON the Multi Gas Monitor. You need to let the Multi Gas Monitor –

INNOVA 1316A-3 warm up for at least 5 minutes. For optimum span

calibration accuracy, let the Multi Gas Monitor –INNOVA 1316A-3 warm

up for at least 30 minutes.

While the Multi Gas Monitor –INNOVA 1316A-3 is warming up, perform

the following steps:
1) If a New NO Sensor is required see the Status window

figure 4.13

,

install a new NO sensor Order number VM0201A.

2) If the New O

2

Sensor is required see the Status window

figure 4.14

,

install a new O

2

sensor. Order number VM0200A.