4 the ndir temperature measurements – LumaSense Technologies INNOVA 1316A-3 User Manual

Page 14

Advertising
background image

Chapter 1
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 14 of 96

µm (2260 cm

-1

), and the filter associated with Carbon Monoxide is

centred on 4.65 µm (2150 cm

-1

).

In the measurement chamber the three beams, which differ in

wavelength frequency, each excite their respective gas (if present).

The absorption of the incident light will be registered by the

Detector.

System Board Assembly

The system board assembly is a surface-mount printed circuit board

that includes the digital and analogue electronic components and the

IR detector assembly. The IR detector assembly includes thermopile

IR detectors for the Ethanol, CO, CO

2

, and reference channels. Light

passed through the band pass filters is directed through the

thermopile window to a detector substrate. The substrate has a light-

sensitive coating, which produces a voltage that is proportional to

light intensity. A reference voltage is produced by the reference

detector, which provides an output that varies with light intensity

and is not affected by the sample gas. The Ethanol detector includes

a thermistor. The detector block temperature measurement

facilitates compensation for IR filter and IR detector temperature

variances. The detector outputs are amplified and sent through a

multiplexer to an ADC converter. The microprocessor samples the

ADC converter output at 200 Hz and derives an uncompensated

digital value representing each gas concentration. The data are

subjected to a compensation routine that provides adjustment for

temperature and pressure variances. The digital value representing

each gas concentration is derived.

1.4.4 The NDIR Temperature Measurements.


The analyzer incorporates several thermistors for the measurement

of temperatures.

IR Detector/Filter Temperature (Ethanol Temperature):

A thermistor located in the optical assembly measures the IR

detector and IR filter temperature. It is used as follows:
“Gas law compensation” and detector response compensation during

Ethanol, CO and CO

2

gas concentration measurement.

Span calibration (via the use of temperature compensated Ethanol,

CO

2

and CO data).

An “out of range” IR detector/filter temperature condition causes

Ethanol, CO

2

and CO data to be declared “data invalid.”