Chapter 5
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 95 of 96

5.1.8 O

2

and NO Sensor Replacement

The optional O

2

and NO sensors have an expected life of 18 months under

normal operating conditions. Follow the procedure defined below to replace

these sensors.
a) Verify that main power turned OFF.
b) Remove the cover.

2) Sensor Removal:

a) Each sensor includes a pull tab. Be careful not to tear it.
b) Carefully pull on the tab and remove the sensor from the 1316A-3 Multi

Gas Monitor UA1372 manifold.

3) Sensor Installation:

a) Position the new sensor at the sensor manifold. Ensure that the sensor’s

connector pins are aligned.

b) Carefully push on the sensor to seat it in the Manifold UA1372.
c) Replace the cover.

4) Sensor Calibration: Refer to

Chapter 4

.

5) Verification: Consult the Status window and verify that no errors are

reported.


5.1.9

Field replaceable parts.

LumaSense Part number Description

DS0790

Zero ref. filter

DS2306

Inlet filter

VF0007A

Fuse 1.6A T

VM0200A

O

2

electrochemical

Sensor

VM0201A

NO electrochemical

Sensor

UA1372A

Mounting kit for

electrochemical Sensors

UH5040A

Pump

UM1118A

Tree way valve

ZG0370A

Power supply