Pvcheck – HT instruments PVCHECK User Manual
Page 15
PVCHECK
EN - 13
5.2.
EFF – EFFICIENCY TEST SETTINGS FOR PV INSTALLATIONS
The purpose of this measurement is to evaluate the DC efficiency of a photovoltaic
installation, with the possibility of obtaining a positive or negative result of the
check/recording depending on a limit on the nDC parameter, freely set by the user. For
this test, the use of the optional remote unit SOLAR-02 is necessary (see § 6.1)
5.2.1. Instrument
settings
1. Position the cursor onto EFF by using the arrow keys
(
,) and confirm with ENTER. The display shows the
values of the output electrical parameters of the
photovoltaic generator.
15/05/12 15:34:26
I r r
- - -
W / m 2
P n o m
3 . 5 0 0
k W
T c
- - -
° C
T e
- - -
° C
P d c
0 . 0
k W
V d c
0 . 0 0 0
V
I d c
0 . 0
A
n d c
- - -
G O t o S t a r t
Selection
E F F
2. Press the ENTER key. The instrument shows the
following options: System parameters and Instrument
settings.
3. Use the arrow keys (
,) to select “Instrument
settings” and confirm with ENTER. The instrument shows
the following screen:
15/05/12 15:34:26
I r r
- - -
W / m 2
P n o m
3 . 5 0 0
k W
T c
- - -
° C
T e
- - -
° C
P d c
0 . 0
k W
V d c
0 . 0 0 0
V
I d c
0 . 0
A
n d c
- - -
System parameters
I n s t r u m e n t s e t t i n g s
Selection
E F F
4. By using the arrow keys (
, ) it is possible to set:
the integration period (IP) which can be used by the
instrument when testing the parameters of a PV
system. The following values can be selected: 5s, 10s,
30s, 60s, 120s, 300s, 600s, 900s, 1800s, 3600s.
The FS of the DC clamp used for measuring DC
current with value selectable between 1A
3000A
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
15/05/12 15:34:26
I P
: 5 s
F S D C c l a m p
: 1 0 0 0 A
S A V E t o s t o r e
E F F