Pvcheck, Measuring… selection m, Outcome: ok m – HT instruments PVCHECK User Manual

Page 46

Advertising
background image

PVCHECK

EN - 44

4. Press GO/STOP to start the test. In case no error

conditions occur, the instrument displays the message
Measuring…” as shown in the screen to the side.

15/05/12 15:34:26

I n s . T e s t

:

1 0 0 0

V

R i m i n

:

1 . 0

M

M o d e :

S t r i n g a

V t e s t =

1 0 2 0

V

R p

=

- - -

M

Measuring…

Selection

M

5. When measurement is complete, the instrument provides

the Ri value, i.e. the minimum value of insulation
resistance of the PV module/string (or of other objects)
being tested, continuously measured during the whole
duration of measurement. If the result is higher than the
minimum limit set, the instrument displays the message
Outcome:OK”; otherwise, it displays the message
Outcome:NO” as shown in the screen to the side.

6. Press the SAVE key to store the test result in the

instrument’s memory (see § 7.2) or the ESC/MENU key
to exit the screen without saving and go back to the main
measuring screen.





















15/05/12 15:34:26

I n s . T e s t

:

1 0 0 0

V

R i m i n

:

1 . 0

M

M o d e :

S t r i n g a

V t e s t =

1 0 2 0

V

R p =

> 1 0 0

M

Outcome: OK

M






Advertising