Pvcheck – HT instruments PVCHECK User Manual

Page 51

Advertising
background image

PVCHECK

EN - 49

6.5.3.1. Anomalous

situations

1. In case the instrument detects a voltage higher than 5V at

its terminals E and C, it does not carry out the test, gives
out a long sound and displays the message “Voltage >
Lim”.

15/05/12 15:34:26

R P E m a x

:

1

R c a l

:

- - -

R p e

=

- - -

I t e s t

=

- - -

m A

V o l t a g e > L i m

Selection

L O W


2. In case the instrument detects that the calibrated

resistance is higher that the measured resistance, the
instrument gives out a long sound and displays the
message: “Calibration not OK”

15/05/12 15:34:26

R P E m a x

:

1

R c a l

:

- - -

R p e

=

- - -

I t e s t

=

- - -

m A

C a l i b r a t i o n n o t O K

Selection

L O W


3. In case the instrument detects a resistance higher than

5

 at its terminals, it gives out a long sound, zeroes the

compensated value and displays the message
“Calibration reset”.

15/05/12 15:34:26

R P E m a x

:

1

R c a l :

0 . 0 0

R p e

=

- - -

I t e s t

=

- - -

m A

C a l i b r a t i o n r e s e t

Selection

L O W