LIST OF DISPLAY MESSAGES

MESSAGE DESCRIPTION

Function not available

The selected function/characteristic is not available

Data not saved

The instrument was not able to save the data

Wrong date

Set a correct system date

RADIO transmission error

The instrument does not communicate via RF with external units

SOLAR-02: Firmware incorrect

Inconsistent SOLAR-02 FW. Update firmware.

Firmware incorrect

Unsuitable instrument FW. Update firmware.

Error 4: contact service

Internal instrument error

Database full

The number of modules added to the internal DB is > 30

Module already added

The name of the module has already been added to the DB

Memory full

Instrument memory full upon pressing the GO key

Error: Vmpp >= Voc

Check module settings inside DB

Error: Impp >= Isc

Check module settings inside DB

Error: Vmpp * Impp >= Pmax

Check module settings inside DB

Error: Alpha too high

Check module settings inside DB

Error: Beta too high

Check module settings inside DB

Error: Gamma too high

Check module settings inside DB

Error: Toll too high

Check module settings inside DB

Wait data analysis

Download data from SOLAR-02 and wait for the result of PV efficiency

test

Download error

Contact service department

Error while saving

Problems with the access to memory area

Remote unit undetected

The instrument does not detect any remote unit SOLAR-02

Unable to perform analysis

Problems with the data downloaded from SOLAR-02. Check settings.

Data unavailable

Generic error. Repeat the test.

Negative voltage

Check the polarity of the instrument’s input terminals

Low voltage

Check the voltage between input terminals P and N

Vin > 1000

Voltage between input terminals > 1000V

Wrong no. of modules. Continue?

Set no. of modules inconsistent with measured Voc

Ref. cell temp. exceeding limits

Temperature measured by reference cell too high

Cell temp. undetected

(ENTER/ESC)

Measurement not carried out on module cell

Low battery

Low battery charge. Insert new batteries.

Please wait for cooling…

Instrument overheated. Wait before resuming the test.

Irradiance too low

Irradiance value lower than the minimum threshold set

NTC Error

Internal NTC efficiency jeopardized. Contact service department

Isc current too high

Measured Isc current > 10A

Current < Lim

Current between P and N lower than the minimum detectable value

EEPROM error: contact service

Internal instrument error

FRAM error: contact service

Internal instrument error

RTC Error: contact service

Internal instrument error

RADIO error: contact service

Internal instrument error

FLASH error: contact service

Internal instrument error

IO EXP error: contact service

Internal instrument error

Voltage > limit

Voltage between terminals E and C > 10V

Label already assigned

Change the numeric reference of the marker associated with the measure

Isc current < Lim

Isc current lower than the minimum detectable value. Contact service

department

CAUTION: internal short-circuit

Contact service department

CAUTION: blown fuse

Contact service department

Reset calibration. Press ENTER.

Resistance value of input cables > 2

Calibration not OK

Calibrated resistance value > measured resistance

Error: Isc offset measurement

Internal instrument error

Rcal > measured R

Calibrated resistance value > measured resistance

CAUTION: AC voltage at P-N

terminals

Presence of AC input voltage

Wait for condenser discharge

Wait for discharge of tested object after insulation