Fig. 8: Connection for IVCK test with direct irradiance measurement

CAPTION:

E: Green

cable

C: Blue

cable

P: Red

cable

N: Black cable

1. PV module/string
2. Main system earthing
3. Earthed metal structure in the

system

4. Reference cell for irradiance

measurement

5. Temperature

sensor

(if

required)

6. Unità

remota

SOLAR-

02Remote unit SOLAR-02

Fig. 9: Connection for IVCK test with irradiance measurement through SOLAR-02


14. In the initial screen of IVCK mode, the following values

are displayed in real time:

Module  type of module being tested
Vdc  value of output voltage from module/string
Irr  irradiance (from a direct measurement or

SOLAR-02 connected via RF)

Tc  modulo temperature (in MAN or AUX mode) and

the relevant measuring mode, or “- - -” in AUTO mode

 In case, the symbol “

” of RF connection with

SOLAR-02

15/05/12 15:34:26

M o d u l e :

S U N P W R 3 1 8

V d c :

5 4 8 . 0

V

I r r

:

8 5 6 W / m 2

T c

:

A u t o ° C

V o c , I s c :

R i ( 1 0 0 0 V )

- - -

M

R p e ( C a l )

- - -

Selezione

I V C K

CAUTION

Upon pressing the GO/STOP key, different error messages can be displayed
by the instrument (see § 6.6) and, therefore, the test cannot be started. Check
and eliminate, if possible, the problem causing the error message before going
on with the test.

15. Press GO/STOP to start the test. In case no error

conditions occur, the instrument displays the message
Measuring…” and the measure of open-circuit voltage
between the terminals P and N and of short-circuit current
(for Isc values

10A).

15/05/12 15:34:26

M o d u l e :

S U N P W R 3 1 8

V d c :

5 4 8 . 0

V

I r r

:

8 5 6 W / m 2

T c :

A u t o

° C

V o c , I s c :

R i ( 1 0 0 0 V )

- - -

M

R p e ( C a l )

- - -

Measuring…

Selection

I V C K