HT instruments PVCHECK User Manual

Page 27

Advertising
background image

PVCHECK

EN - 25

15. Upon reaching “00” after pressing the GO/STOP key, the

test is started and the two units are synchronized with
each other. In these conditions the message “Recording
running…
” is displayed on the main unit and
Recording…” is displayed by SOLAR-02.

15/05/12 15:35:00

I r r

- - -

W / m 2

P n o m 3 . 5 0 0

k W

T c

4 5

° C

T e

3 0

° C

P d c

3 . 1 2 5

k W

V d c

3 8 9

V

I d c

8 . 0 1

A

n d c

- - -

R e c o r d i n g r u n n i n g …

Selection

E F F

16.

At any time it will be possible to analyze the current
recording status by pressing the MENU key. The following
information will be shown:
 Starting date and time of recording
 The value set for the integration period (IP)
 The number of periods elapsed from the beginning of the

recording

 The remaining memory capacity for recording

Press the ESC key to exit the screen

15/05/12 15:35:00

S t a r t 1 5 / 0 5 / 1 2 1 5 : 3 0 : 0 0
P e r i o d : 5 s

I P N u m b e r 6 1

R e c . T i m e

0 d 1 h

R e c o r d i n g r u n n i n g …

17. Now it is possible to bring the unit SOLAR-02 near the PV strings to measure irradiation and

temperature by means of the relevant probes. When the distance between the unit SOLAR-
02 and PVCHECK does not allow a RF connection, on the display of SOLAR-02, the symbol

” flashes for approx. 30s and then disappears, while PVCHECK continues searching

for the connection for approx. 1 minute.

18. Position the reference cell onto the surface of the PV modules. Please refer to the relevant

User Manual for a correct assembly.

19. Put the temperature sensor in contact with the rear side of the module and fasten it with

some tape; avoid touching it with your fingers in order not to alter the measurement result.

20. Wait for a few seconds to allow the probes to reach a steady measure and then connect the

irradiation probe to input PYRA/CELL and the temperature probe to input TEMP of unit
SOLAR-02.

21. Wait for the message “READY” to appear on the display of SOLAR-02 to indicate that the

unit has detected the data with solar irradiation > minimum threshold set (see § 5.1.5).

22. With the message “READY” shown on the display, wait for approximately 1 minute in

order to take a certain number of samples.

23. Disconnect the irradiation and temperature probes from unit SOLAR-02 and bring the unit

near PVCHECK (max distance 1m).

24. The main unit PVCHECK must be in EFF mode. If there is no flashing symbol “

”,

press key

 to activate the RF connection search again.

25. Press key

 on SOLAR-02 to activate the RF connection again. Consequently, the main

unit will show the message “active radio connection”.

26. To stop testing, press the GO/STOP key on the instrument and confirm with ENTER that

you want to stop recording.

27. The display of the unit SOLAR-02 will display the message “SEND” to indicate that the data

will be transferred to the main unit.