KYORITSU 6050 User Manual

Page 10

Advertising
background image

̶ 8 ̶

● UL value selector

Select UL (limit of contact voltage value) 25V

or 50V. Where Uc (contact voltage) exceeds

UL value at RCD testing, "Uc H" will be

displayed without starting the measurement.

● Voltage Indication

When connecting Test Lead to circuit, voltage

between L-PE is displayed. If the voltage is

lower than 100V, displayed as "V L-PE Lo".

And if it is higher than 260V, displayed as "V L-

PE Hi".But on L-L measurement range,

displayed as "V L-PE Hi" if the voltage is 440V

or more.

● Low battery

  warning

" "symbol appears in the display if the battery

voltage drops below 8V.

● Auto power off

Automatically switches the instrument off after

a period of approximately 10 minutes.

● Display

The liquid crystal display has 3 1/2 digits with a

decimal point and units of measurement (V,

Ω,A,kA,mA,ms )

● Data memory,

  Communication

  facility

Can store 300 measured result.

Data can be transferred to PC via Optical

Adapter Model 8212.

● Optional

  Accessory

1) Model 7121 for measuring distribution board

or lighting circuit test lead.

2) Optical Adapter Model 8212 (with PC

software "Kew Report")

● Auto Test

Instrument becomes Auto test mode by turning

Test button to right and locked it. Under Auto

test mode, test starts automatically only by

connecting Test Lead to a measuring circuit.