Personal Ambulatory Oxygen System Provider Technical Manual

10

PN 20631679 Rev C

Guidance and manufacturer’s declaration–electromagnetic immunity

The Eclipse 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse 5

should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electromagnetic

environment –

guidance

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic

tile. If floors are covered with synthetic mate-

rial, the relative humidity should be at least

30 %.

Electrical fast tran-

sient/burst

IEC 61000-4-4

±2 kV for power supply

lines

±1 kV for input/output lines

±2 kV for power supply

lines

N/A

Mains power quality should be that of a typi-

cal commercial or hospital environment.

Surge

IEC 61000-4-5

±1 kV line(s) to line(s)

±2 kV line(s) to earth

±1 kV line(s) to line(s)

±2 kV line(s) to earth

Mains power quality should be that of a typi-

cal commercial or hospital environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines IEC

61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be that of a typi-

cal commercial or hospital environment. If

the user of the Eclipse 5 requires continued

operation during power mains interruptions,

it is recommended that the Eclipse 5 be pow-

ered from an uninterruptible power supply or

a battery.

Power frequency

(50/60 Hz) mag-

netic field IEC

61000-4-8

3 A / m

3 A / m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a

typical commercial or hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.