Personal Ambulatory Oxygen System Provider Technical Manual

65

PN 20631679 Rev C

Figure 35: Tube Routing under ATF Module.

10. Place the ATF in the Unit Case and remove the caps from the ATF Ports. Install the Braided Tubes into

the ATF pressure and vacuum ports as shown on Figure 32. Do not use oil or grease if the tubing is
difficult to install. Secure joints with cable ties.

11. Mount the ATF with 3 M4x16 screws and 3 washers as shown on Figure 33 on previous page. Apply

Loctite

®

425 Thread Locker on the threaded tips of the screws before installing. Tighten screws but DO

NOT squash the grommets (2 in.lbs. maximum).

If needed, install spacer (20623336) underneath the ATF mount closest to the product tank with a longer

screw (6961-418-SEQ) to ensure that the ATF is level.

NOTE:

Inspect all tubing before securing with a cable tie.

CAUTION

Remove caps on ATF ports ONLY before connecting the tubes. DO NOT
leave the ports open unless ready to install the tubes. Also place the
removed caps from the new ATF and place on the ports of the replaced
ATF.

DO NOT use any petroleum based or other lubricants. A spontaneous
and violent ignition may occur if oil, grease or other petroleum
substances come into contact with oxygen under pressure. Keep these
substances away from the oxygen system, tubing and connections
and any other oxygen source.

WARNING:

16” Silicone Tube

Routing

NOTE:

Over tightening may cause excess vibration.