Personal Ambulatory Oxygen System Provider Technical Manual

52

PN 20631679 Rev C

ELECTRICAL SAFETy TEST

This is required only for the Eclipse 5 Oxygen System, that is used in a hospital or institutional environment. This is not required
for home care use.

To test the basic electrical safety of the Eclipse 5 AC Power Supply, SeQual recommends using an LKG-601 Electrical Safety
Analyzer (Netech Corporation, Hicksville, NY) or equivalent to verify that the current leakage to ground is within appropriate
limits.

1. Plug the AC Power Supply into the electrical safety analyzer. Disconnect the AC Power Supply from the Eclipse

5 unit.

2. Plug the electrical safety analyzer into a wall outlet.

3. Follow the analyzer manufacturer’s instructions for measuring both the forward and reverse earth leakage

current only. Verify that forward and reverse-current leakage to ground is < 250 µA (100/115VAC applications)
and < 500 µA (220/240VAC applications).

4. Disconnect the AC Power Supply from the electrical safety analyzer.

5. If the AC Power Supply measures leakage current greater than the criteria in step 3, please return it to SeQual

for service.

6. EDAT may be utilized to record Eclipse 5 device status and dates of service.

CLEANING THE ECLIPSE 5

Clean inside the unit, as needed, using a small vacuum cleaner or brush to remove any accumulation of dust or debris prior to
attaching the covers. After reinstalling the cover, verify that the rubber gasket is installed correctly.

Use mild detergent solution to clean the cabinet, Control Panel and power supplies.

Turn OFF the Eclipse 5 and discon nect

from AC or DC power before any cleaning or disinfection activity.

DO NOT

spray the cabinet Control Panel or power supplies.

Use a damp (not soaking wet) cloth or sponge. Spray the cloth or sponge with a mild detergent solution to clean the cabinet
and power supplies. To disinfect the Eclipse 5, use Lysol

®

Brand II disinfectant. Proceed as directed by the manufacturer.

Unplug Power Cords, AC or DC Power Supplies before cleaning the
exterior cabinet. DO NOT use denatured alcohol or apply liquid
spray or aerosol cleaners.

WARNING: